Delta reduces leakage current to increase user safety for MEB-1K5A24T AAA series medical power supplies. Please refer to the table below for information on the changes. Comprehensive testing and verification of the results have been conducted to ensure that all customers can expect consistent, high-quality product performance after these design changes.
If you require any further clarification, please refer to the latest specifications on the DeltaPSU website or reach out to us via our Contact Us Form.
Model List:
| Model Name | Description | Old Revision |
New Revision |
Implementation Date |
| MEB-1K5A24T AAA | Reduce leakage current | 00 | 01 | 2025/10/16 |
Modification of MEB-1K5A24T AAA:
| Specifications Before the Change | Specifications After the Change |
|
| Touch Current (maximum) | 0.1 mA @ Normal Condition 0.5 mA @ Single Fault Condition |
0.1 mA @ Normal Condition 0.2 mA @ Single Fault Condition |
| Earth Leakage Current (maximum) | 0.5 mA @ Normal Condition 1.0 mA @ Single Fault Condition |
0.2 mA @ Normal Condition 0.4 mA @ Single Fault Condition |